Audi Q3 Owners & Service Manuals

Audi Q3: Lane Change Assistance, Calibrating

Special tools and workshop equipment required

  • Calibration Tool -VAS6350-

Conditions

- The Lane Change Assistance Control Module -J769-/ Lane Change Assistance Control Module 2 -J770- must be calibrated using the Vehicle Diagnostic Tester during the following conditions:

  • Lane Change Assistance Control Module -J769- or Lane Change Assistance Control Module 2 -J770- was replaced.
  • The rear bumper cover was damaged.
  • The rear bumper cover was removed and installed.
  • The DTC "no or incorrect basic setting/adaptation" is stored in the DTC memory.

Preparing for Calibration

Audi Q3. Preparing for Calibration

1 - Calibration Tool - Spacing Laser -VAS6350/2-

  • For distance measurement
  • For information on usage. Refer to the Operating Instructions.

2 - Calibration Tool -VAS6350-

3 - Level

  • On the Calibration Tool -VAS6350-
  • To check the horizontal position of the Calibration Tool -VAS6350-

4 - Catch Bracket

  • To mount the Calibration Tool - Spacing Laser -VAS6350/2- for the distance measurement
  • Distance to the Calibration Tool - Wheel Center Mountings -VAS6350/1- on the rear wheels: dimension -a- = 1700 +- 2 mm

5 - Calibration Tool - Wheel Center Mountings -VAS6350/1-

  • With wheel bolt adapter and measuring paddle

6 - Brand Emblem

  • The laser point is aligned on the center of the brand emblem

7 - Calibration Tool - Lane Change Calibration Tool -VAS6350/4-

  • Is moved from one side of measuring field to the other during calibration
  • When installed correctly, vehicle electrical system voltage line must be connected at bottom left of calibration device (as seen in direction of travel)

8 - Measurement Scale

  • For positioning the Calibration Tool - Lane Change Calibration Tool -VAS6350/4-

9 - Calibration Tool - Linear Laser -VAS6350/3-

  • With "laser eye protection"
  • On the Calibration Tool -VAS6350-
  • To turn on and off. Refer to the Calibration Tool -VAS6350- Operating Instructions.

10 - Plastic Foot

  • Quantity: 3
  • Can be adjusted for adjusting the horizontal position of the Calibration Tool -VAS6350-

Procedure

- Move the vehicle onto a solid, flat surface.

- Set the parking brake - the vehicle must not move during the measuring.

- Place the front wheels in a straight-ahead position - steering wheel in 0 position.

- Connect the vehicle diagnostic tester. Refer to Vehicle Diagnostic Tester.

  Note

If a malfunction message appears in the display. Refer to Vehicle Diagnostic Tester.

- Turn on the ignition.

- Remove sticker with metal foil from bumper cover if necessary.

- Secure three suitable wheel bolt adapters for the wheel bolts on each Calibration Tool - Wheel Center Mountings -VAS6350/1-.

- Insert the measuring paddle on both Calibration Tool - Wheel Center Mountings -VAS6350/1- and secure with lock nut.

- Place the Calibration Tool - Wheel Center Mountings -VAS6350/1- onto the wheel bolts on both rear wheels.

  • The wheel center sensor rotation center must be in wheel rotation center.

  Note

Place the Calibration Tool - Wheel Center Mountings -VAS6350/1- on wheels so that "anti-theft wheel bolts" are not connected with wheel center mounting.

- Adjust the measuring paddle with aid of locking nuts so that they move freely just above the floor.

  • The measuring paddles must move easily.
  • The measuring paddles must be vertical.

Audi Q3. N94-10685

- Position the Calibration Tool -VAS6350- at distance -a- from rear wheels.

  • Dimension -a- = 1700 mm.

Audi Q3. A96-10286

- Switch on the Calibration Tool - Spacing Laser -VAS6350/2- with the ON button.

Display on the Calibration Tool - Spacing Laser -VAS6350/2-:

  • "- - - m"

  Note

The laser is switched on at same time.

Audi Q3. N94-10681

- For distance measurement, hold the Calibration Tool - Spacing Laser -VAS6350/2--2- flush to the right stop bracket as illustrated.

  • The Calibration Tool - Spacing Laser -VAS6350/2- must lie firmly against the catch bracket.

- Make sure the "laser beam" for the distance measurement contacts the paddle on the lower enlarged part -1-.

If this is not the case, measure paddle height must be corrected using locking nuts on Calibration Tool - Wheel Center Mountings -VAS6350/1-.

Audi Q3. N94-10684

- Briefly press the ON button for distance measurement.

Display on the Calibration Tool - Spacing Laser -VAS6350/2-:

  • "1.700 m" (specified value: 1700 +- 2 mm).

- Repeat measurement procedure from left catch bracket to left rear wheel.

  • The distance value must be the same on both sides.

If both measured values are not the same, adjust the Calibration Tool -VAS6350- accordingly.

Performing Calibration

Vehicle Diagnostic Tester is attached.

- Choose operation mode Diagnosis and start the diagnosis.

- Select the tab test plan.

- Select select individual tests and choose the following sequence.

  • Body
  • Electrical Equipment
  • 01 - OBD-capable systems
  • 3C - Lane change assistance control module - J769
  • 3C - Lane change assistance control module, functions
  • 3C - Calibration

The Vehicle Diagnostic Tester continues with the calibrating procedure from here on.

  Note

  • The following information clarifies which version of the lane change assistance is installed (version SWA1.7).
  • This information for adjusting the calibration device is important for later steps in the procedure to avoid malfunctions.

Audi Q3. A96-10289

- Secure the Calibration Tool - Lane Change Calibration Tool -VAS6350/4- at right rear of the Calibration Tool -VAS6350- mount.

  • When installed correctly, vehicle electrical system voltage line must be connected at bottom left of calibration device (as seen in direction of travel).

Dimension -a- is measured from the upper edge of calibration device to the floor.

Audi Q3. A96-10625

Audi Q3. Performing Calibration

- The indicator on the base of the Calibration Tool - Lane Change Calibration Tool -VAS6350/4- must align with adjustment dimension on the measuring scale -1- on measuring field -arrow-.

Audi Q3. Performing Calibration

- Connect the Calibration Tool - Lane Change Calibration Tool -VAS6350/4- to the vehicle electrical system voltage.

- Bring Calibration Device -VAS6350- into horizontal position by turning plastic base.

  • Observe the bubble level (sight glass) on the Calibration Tool -VAS6350--arrow-.

Audi Q3. N94-10686

- Switch on the Calibration Tool - Linear Laser -VAS6350/3- on the Calibration Tool -VAS6350-.

- Put on "Laser eye protection".

- Align the Calibration Tool -VAS6350- so that the Calibration Tool - Linear Laser -VAS6350/3- strikes the center of the brand emblem on the rear lid.

Audi Q3. A96-11039

- Check the left and right distance between the catch bracket on the Calibration Tool -VAS6350- and measuring paddle -1- on wheel mountings again.

  • Specified value: 1700 +- 2 mm

Calibration Procedure

The following should not occur during the calibration procedure:

  • No metal reflectors may be located within a two m radius of the calibration device (for example, tool carts, metal cabinets).
  • Vehicle doors must not be opened or closed.
  • People should not sit in the vehicle.
  • People must not go between the vehicle and the Calibration Tool - Lane Change Calibration Tool -VAS6350/4-.

Audi Q3. N94-10684

Procedure

- Switch on the Calibration Tool - Lane Change Calibration Tool -VAS6350/4- at the power switch -3-.

  • The green LED -1- must light up.

  Note

If the red LED -2- illuminates: Check the Calibration Tool - Lane Change Calibration Tool -VAS6350/4-.

- Follow the instructions on the Vehicle Diagnostic Tester display.

Audi Q3. A96-10358

During the course of the program, a request is made to switch the calibration unit for lane change assist -VAS6350/4- from the right to the left side of the Calibration Tool -VAS6350-.

- Switch the Calibration Tool - Lane Change Calibration Tool -VAS6350/4- off and move device.

  • When installed correctly, vehicle electrical system voltage line must be connected at bottom left of calibration device (as seen in direction of travel).

Dimension -a- is measured from the upper edge of calibration device to the floor.

Audi Q3. A96-10626

Audi Q3. Calibration Procedure

- The indicator on the base of the Calibration Tool - Lane Change Calibration Tool -VAS6350/4- must align with adjustment dimension on the measuring scale -1- on measuring field -arrow-.

Audi Q3. Calibration Procedure

- Switch on the Calibration Tool - Lane Change Calibration Tool -VAS6350/4- at the power switch -3-.

  • The green LED -1- must light up.

- Follow the instructions on the Vehicle Diagnostic Tester display.

- After calibrating the lane change assistance successfully, end "calibration", switch off the ignition and disconnect the diagnostic connector.

Audi Q3. A96-10358

    READ NEXT:

     Driver Assistance Systems Front Camera

    Driver Assistance Systems Front Camera Component Location Overview 1 - Retaining Plate There are different versions. Refer to the Parts Catalog. With heated windshield Cannot be s

     Trailer Hitch

    Overview - Trailer Hitch Socket and Towing Recognition Control Module 1 - LED Indicator Lamp Trailer Hitch -Locked- Indicator Lamp -K226-, Trailer Hitch -Unlocked- Indicator Lamp -K

     Wiring

    SEE MORE:

     Overview

    Applies to: vehicles with emergency call function An emergency call is a combination of data transmission and a phone call. Your vehicle uses data transmission to forward important information, such as the vehicle and position data, to an emergency call center. Depending on the country and the vehi

     Installation Instructions, Run-Flat Tires

    Warm up cold tires to the minimum mounting temperature   Note This applies also to ultra high performance tires (height-/width ration smaller/same 45% and speed rating symbol larger than/same as V).   WARNING The minimum mounting temperature for a tire may not be below 21℃

    © 2019-2024 Copyright www.auq3.net